Cardiac output (CO) is the amount of blood that is ejected from the heart per minute. Monitoring the components of CO in patients is important to ensure there is sufficient volume of blood in the body to transport oxygen. Monitoring CO can help clinicians manage fluids, monitor therapeutic interventions, and improve patient outcomes.
CO monitoring, along with other advanced haemodynamic parameters, are available through the Masimo LiDCO® Haemodynamic Monitoring System.
Fluid administration is one of the most common interventions made to increase CO. However, fluid administration should be balanced to avoid both hypovolemia and hypervolemia, which have both been associated with negative outcomes.1
Oxygen delivery (DO2) is the amount of oxygen delivered to the tissues, calculated as the product of CO and oxygen content (CaO2).3
CO is calculated by multiplying the stroke volume (SV) by the patient’s heart rate (HR). SV is the amount of blood pumped by the left ventricle of the heart in one contraction.3
DO2 and CaO2 are influenced by the patient's Oxygen Saturation (SaO2) and Haemoglobin (Hb).3
Parameter | Equation | Normal Range |
Oxygen Delivery (DO2) | CaO2 x CO x 10 | 950-1150 mL/min |
---|---|---|
Cardiac Output (CO) | HR x SV/1000 | 4.0 – 8.0 L/min |
Oxygen Content (CaO2) | (0.0138 x Hgb x SaO2) + (0.0031 x PaO2) | 17 – 20 mL/dL |
Stroke Volume (SV) | N/A | 60 – 100 mL/beat |
Oxygen Saturation (SaO2) | N/A | 95 – 100% |
Total Haemoglobin (Hb) | N/A | Men: 13.8-17.2 g/dL Women: 12.1-15.1 g/dL |
Heart Rate (HR) | N/A | Range varies based on patient status (resting vs. active, age, etc.) |
The PulseCO algorithm provides continuous beat-to-beat CO and SV by analysing a blood pressure waveform. The algorithm is based on physics and physiological principles and focuses on pulse power analysis rather than waveform shape or contour. Unlike other arterial pressure algorithms, PulseCO is not based on statistics and assumptions about vascular compliance, nor on the detection of the dicrotic notch, which is often a challenge with peripheral arterial signals. As a result, the PulseCO algorithm avoids the limitations of other pulse pressure or contour-based haemodynamic monitoring technologies.
The current gold standard in haemodynamic monitoring, although not as commonly used due to its invasiveness, is the pulmonary artery catheter (PAC). The PulseCO algorithm has been validated against the PAC, demonstrating a good agreement between the two methods.9
In a study comparing the outcomes of 600 emergency laparotomy patients, researchers found that, following the implementation of a program including LiDCO monitoring with PulseCO technology, there was a significant decrease in mortality at 30 days (from 21.8 to 15.5%) and 180 days (from 29.5 to 22.2%).10
In a randomised, controlled trial of 743 patients undergoing major abdominal surgery, researchers found haemodynamic optimisation with LiDCO monitoring with PulseCO technology led to a 20% reduction in postoperative complications and, as a result, patients monitored with LiDCO Monitoring with PulseCO technology were on average $530 less expensive to treat than control patients who were not monitored.1
References:
Pearse R et al. JAMA 2014; 311(21):2181-90.
Bellamy MC. Br J Anaesth. 2006 Dec;97(6):755-7.
Miller's Anesthesia, 8th Edition, Vol. 2
Burns, S. M., & Delgado, S. A. (2019). AACN essentials of critical care nursing (4th ed.). New York, NY: McGraw-Hill.
Diepenbrock, N. H. (2015). Quick reference to critical care (5th ed.). Philadelphia, PA: Wolters Kluwer.
Jones, J., & Fix, B. (2015). Critical care notes: Clinical pocket guide (2nd ed.). Philadelphia, PA: FA Davis.
Urden, L. D., Stacy, K. M., & Lough, M. E. (2020). Priorities in critical care nursing (8th ed.). St. Louis, MO: Elsevier.
World Health Organization Global Database on Anaemia. 2008.
Costa et al. Intens Care Med. 2007. DOI 10.1007/s00134-007-0878-6 P1.8
Tengberg LT et al. Br J Surg 2017; 104:463-471.